The installation of the grating ruler displacement sensor reading head
When installing the reading head, you should first ensure that the base of the reading head meets the installation requirements and then installs the reading head. The mounting method is similar to the master. Finally adjust the reading head, so that the reading head and grating main square parallel to ensure that within 0.1mm, the reading head and the main ruler of the gap control in 1 ~ 1.5mm or less.